SPECIFICATIONS
Accuracy:
± .05% of reading ± 2 counts of A/D @ 25°C. Temperature coefficient of 25PPM/degree Celsius.
Resolution of A/D:
12 bits
Input bias current:
± 50nA maximum ± 100pA/°C.
NMRR:
70db
CMRR:
70db
Rollover error:
± 3 counts of A/D
Front Panel Controls: (Located behind lens)
Span adjust switches
Tare pushbutton (optional through lens)
Decimal point selector switch (5 decimal points I.X.X.X.X.X)
Digital filter average set: 1 to 100 conversions
Peak and valley reading display and reset
Visual indicators:
5 l/2 digits. .56" red LED with polarity
Outputs: +5V at limit, Four limit outputs, will sink 24mA
ASCII RS232 compatible:
(optional)
ASCII RS422 compatible:
(optional)
Baud Rate:
300,600,1200,2400,4800,9600
Analog output:
(optional) 0 to ± 10VDC @ 1mA max.
Excitation output:
± 15VDC and 5VDC @ 30mA
Sample Rate:
62.5 to 500 per second for peak, valley, and limits.
Display update four times per second.
Input Filter:
2.5 KHz single pole passive filter. Digital filter for display only can be set to be an average of 62.5 to 500 conversions. Front end filter can be set to be an average of 1 to 8 conversions.
Power:
| 115VAC 50-400 Hz | | 8 watts maximum |
| 230VAC 50-400 Hz | | 8 watts maximum (optional)
|
| 10-30VDC 500 mA maximum | | (optional) |
| 5VDC 600 mA maximum | | (optional) |
Remote Inputs:
Tare
Display peak reading
Display valley reading
Remove tare
Reset peak/valley reading
Operating temperature:
-20° to +60° Celsius.
I/O Method:
Input power via screw terminals. 2 subminiature "D" connectors for strain gage hookup and RS232 output. Mating "D" connector for strain gage hookup is supplied with unit. See option 29 for second connector.
|